Status Messages - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

18 T
RANSCUTANEOUS

Status Messages

Message
tpO
> #
2
tpO
< #
2
tpCO
> #
2
tpCO
< #
2
TP Out of Range
(High)
TP Warming Up
-- Reduced
Accuracy.
TP Sensor
Unplugged
TP Pod
Unplugged
Check TP
Sensor Site
TP Check/
Replace Sensor
TP Change
Sensor Site
Remove Probe
from Chamber
TP Chamber
Timeout
18-14
B
LOOD
Possible Cause
tpO
or tpCO
values are outside the
2
2
alarm limits, due to one of the following
conditions:
• Physiological
• Inappropriate limit settings
• Loose fixation ring
tpO
/CO
values are outside the
2
2
measuring range, due to one of the
following conditions:
• Physiological
• Equipment malfunction
Warm-up phase in progress
The tpO
/CO
sensor is unplugged.
2
2
The tpO
/CO
pod is not connected to
2
2
the monitor.
The site timer will expire within 15
minutes.
The auto heater shutdown is in effect or
the sensor didn't heat to the desired
temperature setting.
The site timer has expired.
The calibration of the sensor is
complete.
The electrode remained in the
calibration chamber > 30 min.
D
/D
ELTA
G
M
AS
ONITORING
Suggested Action
• Check the patient and treat, if
necessary.
• Change the alarm limits (see Alarms
chapter).
• Apply a new fixation ring at different
monitoring site.
• Check the sensor and pod;
recalibrate the sensor; change the
sensor membranes or replace the
sensor, if necessary.
• Check the patient and treat, if
necessary.
• Check the sensor and pod;
recalibrate the sensor; apply new
sensor membranes; replace the
sensor, if necessary.
Wait until the warm-up phase has
ended before relying on the displayed
tpO
/CO
2
evaluation.
• Check all connections.
• Replace the sensor and recalibrate,
if necessary.
• Check all connections.
• Replace the pod or connecting cable,
if necessary.
Prepare to recalibrate and reposition
the sensor.
• Recalibrate sensor and apply at new
monitoring site to start the Site
Timer.
• Verify that the sensor is compatible
with pod.
• Immediately remove the sensor from
the patient.
• Recalibrate the sensor and apply at
new monitoring site.
Remove the sensor from the
calibration chamber.
Recalibrate or cover the sensor with a
protective cap for storage.
XL/K
ELTA
APPA
values for clinical
2
VF8

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity delta xlInfinity kappa

Table of Contents