Continuous Recordings; Events And Trends; Pending Recordings - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Continuous Recordings

Unlike timed recordings, which run only for a specified time, continuous recordings
run until you stop them manually.
To request a continuous recording
1. Press the Menu fixed key to open the Main menu.
2. Click on
To stop the recording:
Click on

Events and Trends

The monitor can store waveforms and parameter values for up to 50 events (parameter
alarms, arrhythmia events, marked events). These are displayed on the Event Recall
screen (see page 1-27).
To request Event and Trend Recordings
1. Press the
2. Click on
3. Click on

Pending Recordings

Recorders connected to the monitor may be temporarily unavailable to print (for
example, during a paper change). If another recorder is available, the recording is
rerouted to that recorder and printed there in its entirety. If no recorder is available, the
data becomes a pending recording and is printed as soon as a recorder is available. The
monitor can store up to six timed recordings and one request for a continuous
recording. The print order is determined by the type of recording. Continuous
recordings have the highest priority, followed by timed, and then alarm recordings.
NOTE: When the monitor stores a timed recording, it saves the actual monitoring data
at the time of the recording request. For continuous recordings, however, the monitor
saves only the recording request, and not the actual data.
VF8
D
.
Cont. Record
again or press the recorder's
Cont. Record
fixed key.
Fast Access
,
Trend Graphs
Trend Table,
at the bottom of the displayed screen.
Record
/D
XL/K
ELTA
ELTA
Stop
or
Event Recall
APPA
R
ECORDINGS
fixed key.
.
7-7