Dräger Infinity Delta Instructions For Use Manual page 348

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

19
CO2 (E
ET
ND
Message
Condition
etCO
Adapter
The airway adapter calibration
2
Cal. Failed
(zeroing in room air) failed.
etCO2
The sidestream measurement
Adapter Cal.
mode was initiated, requiring air
Required
way adapter cal. (zeroing in
room air) to calibrate pump.
etCO
Tubing
Sidestream tubing obstructed, or
2
Blocked
the filter is clogged.
etCO
Tubing
The sidestream tubing has a
2
Leak
leak.
etCO
module
The etCO
2
Disconnected
connected to the monitor.
etCO
module
Corrupt EEPROM. Wrong
2
Incompatible
software or hardware version
NOTE: If the etCO
19-18
-T
CO2)
IDAL
module is not
2
pod is disconnected, the etCO
2
D
/D
ELTA
MONITORING
Suggested Action
Make sure the adapter is properly attached to
the sensor and that its windows are clean. If the
problem persists, try another adapter.
Calibrate the new adapter.
• Clear blockage in the tubing.
• Replace the module/pod.
Change the tubing.
Check cable and connection. Replace cable if
necessary.
• Try a new module.
• Consult the hospital's technical personnel.
parameter box disappears.
2
XL/K
ELTA
APPA
VF8