Message
Possible Cause
RESP Lead Off The cause could be one or more
of the following:
• Broken cable
• Loose lead wire
• Faulty lead wire
• Dried out gel on electrodes
• M
ULTIMED
RESP
The user has turned on
Relearning
respiration monitoring or has
clicked on Relearn.
RESP Artifact
Persistent artifact
RESP H/W
Not applicable
Failure
VF8
D
pod defective
/D
XL/K
ELTA
ELTA
Suggested Action
• Check the patient cable and lead wires
carefully.
• Replace any cable or lead wire that is
suspect.
• Reapply gel or change the electrode.
• Check the M
ULTIMED
necessary.
None required
• Check the patient cable and lead wires
carefully.
• Check the electrode placement. Change
their position, if necessary.
Call your hospital's technical personnel or
DrägerService.
APPA
S
M
TATUS
ESSAGES
pod and replace it, if
12-19