Dräger Infinity Delta Instructions For Use Manual page 389

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

1 Point Calibration (Room Air)
A 1-point calibration of the sensor to room air (21 % oxygen) should be performed on
a daily basis.
1. Make sure the monitor is turned on and that the NeoMed and the monitor are
set up for FiO
2. Expose the sensor to room air.
3. Open the FIO
4. Click on
Calibration may take from 1-10 minutes appears.
5. Wait for the message 21 % Calibration Complete to appear. (A message
informs you of a failed calibration. Try calibrating again; if the message
persists, try a new sensor.)
6. Return the sensor to the incubator or oxygen hood.
2 Point Calibration (Cal Gas)
NOTE: Contact your hospital's technical personnel for help with 2-point calibration.
A 2-point calibration of the system, to 100 % dry oxygen and room air (21 % oxygen),
should be performed every week:
1. Make sure the monitor is turned on, that the NeoMed pod is properly
connected, and set up for FiO
2. Set up the sensor for O
3. Open the FiO
4. Click on
5. Supply 100 % O
6. Click on
in progress and requesting you to wait until calibration is complete before
proceeding with room air calibration.
7. Wait for the system to calibrate. When calibration is finished, the message
100 % Calibration Complete appears.
8. Follow the instructions on page 22-5 to calibrate the system to room air
(1-point calibration).
VF8
D
monitoring (see page 22-3).
2
menu (see page 22-4).
2
The message 21 % Calibration in Progress --
1 Point Cal.
calibration following your hospital's guidelines.
2
menu (see page 22-4).
2
2 Point Cal.
when instructed by the monitor.
2
. A message appears informing you that the calibration is
Continue
/D
XL/K
ELTA
ELTA
monitoring.
2
APPA
F
O2 S
I
ETUP
22-5