Kappa Video Display

The Kappa complies with the requirements of IEC 60601-1 when used with Dräger
approved medical grade displays, available in multiple screen sizes (see page B-18.)
CAUTION: The remote display output on the IDS is not galvanically isolated. If you use
a video monitor other than one specified by Dräger, it must comply with IEC 60601-1
and be suitable for use in the presence of flammable anesthetic mixtures and/or
flammable liquids (see "Safety Considerations" on page 6"). Based on the intended use
of the system, the video monitor must also have suitable classifications for water ingress
protection as well as radiated and conducted emissions. Upon installation, the installer
must make sure that all of these requirements are met.