Selecting A Sensor - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Selecting a Sensor

Select a sensor according to your monitoring needs and the size of your patient's
endotracheal tube. To monitor respiratory mechanics only, use a Flow sensor. For both
respiratory mechanics and etCO
Sensor Type
Flow
(Neonatal)
Flow
(Adult/Pediatric)
Combined CO2/Flow
(Neonatal)
Combined
CO2/Flow
(Pediatric)
Combined
CO2/Flow (Adult/
Pediatric)
To monitor respiratory mechanics only
1. Select an appropriate flow sensor.
2. Plug the sensor tubing into the etCO
3. Connect the flow sensor to the breathing circuit as shown in the following
illustration. A message at the top left of the screen confirms the size of the
sensor and its correct connection.
VF8
D
parameters, use a combined CO2/Flow sensor.
2
ETT Size
< 4 mm
> 4 mm
2.5 - 4 mm
3.5 - 6 mm
5.5 mm
/D
XL/K
ELTA
ELTA
Tidal Volume
1 - 100 ml
200 - 3000 ml
1 - 100 ml
30 - 400 ml
200 - 3000 ml
/Respiratory Mechanics pod.
2
APPA
H
S
ARDWARE
ETUP
Patient Age
Not specified
Not specified
< 4 years
2 - 18 years
> 12 years
21-5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity delta xlInfinity kappa

Table of Contents