Dräger Infinity Delta Instructions For Use Manual page 558

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

B T
D
ECHNICAL
Nellcor Oximax® – SpO2 pod
Physical Attributes
Size (H x W x T)
Weight
Connections
Power source:
Mode of operation:
Degree of protection
against electric shock:
Protection against
ingress of water:
Environmental Requirements
Temperature range:
Relative humidity:
Atmospheric pressure Operating:485 to 795 mmHg (64 to 106 kPa)
B-32
ATA
37 x 58 x 118 mm (1.5 x 2.3 x 4.6 in)
166 g (0.075 lbs)
X8:for power and connection to monitor
Cable connection: for intermediate cable to Nellcor sensors
Powered directly from monitor
Continuous
Type CF
IPX1 (per IEC 60529)
Operating:
0 to 45 °C (32 to 113 °F)
Storage: –40 to 70 °C (–40 to 158 °F)
Operating:10 to 95 %, non-condensing
Storage: 10 to 95 %, (with packaging)
Storage: 375 to 795 mmHg (50 to 106 kPa)
D
/D
ELTA
XL/K
ELTA
APPA
VF8

Hide quick links: