Quick Reference – Arrhythmia Setup Table
Function
Description
Relearn
Initiates a relearn process. Dräger
recommends that you perform a relearn
under the following conditions:
• A lead is reconnected or electrodes are
repositioned.
• Eight hours have passed since last
reference complex learned.
• Questionable arrhythmia calls appear
on the patient's ECG.
• Other significant changes appear on the
patient's ECG.
Alarm
Sets the alarm priority (see page 5-3) for
an arrhythmia event.
NOTE: Settings for Asystole (ASY) and
Ventricular Fibrillation (VF) cannot be
modified.
Rate
Together with the count, the rate
determines the point at which an event
call is triggered
NOTES:
• You cannot modify the rate for the
following parameters: ASY, VF, CPT,
BGM or ARTF.
• RUN and AIVR cannot be modified
because their settings derive from VT.
They are included to quantify their
derivation, based on current VT values.
Count
Together with the rate, the count
determines the point at which an event
call is triggered
NOTES:
• You cannot modify the count for the
following parameters: ASY, VF, CPT,
BGM or ARTF.
• RUN and AIVR derive their settings from
VT and therefore cannot be modified.
They are included in order to quantify
their derivation, based on current VT
values.
9-8
M
ONITORING
D
/D
ELTA
Available Settings
To learn or relearn the template:
1. Set ARR Monitoring to Basic or Full.
2. Verify the quality of the ECG signal.
3. Check that the patient's ECG displays
a normal reference pattern.
4. Click on Relearn to begin a new
learning phase.
• High
• Medium
• Low
(off)
• VT; 100 to 200, increments of 10
• RUN; Same as VT Rate
• AIVR; ≤VT
• SVT; 120 to 200, increments of 10
• TACH; 100 to 200, increments of 10
• BRDY; 30 to 105, increments of 5
• PAUS; 1.0 to 3.5s, increments of 0.5s
• VT; 5 to 15, increments of 1
• RUN; 3 to VT
• AIVR; Count ≥3
• SVT; 3 to 10, increments of 1
• TACH; 5 to 15, increments of 1
• BRDY; not applicable
• PAUS; not applicable
XL/K
ELTA
APPA
- 1
Rate
- 1
Count
VF8