Alarm History Table; Alarm Limits Shortcut; St And Arrhythmia Alarms - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

5 A
LARMS
Alarm Limits Table
Function
Description
Archive
Allows you to store and/or record automatically
an alarm event for the selected parameter.
You can later review stored alarms on the
Event Recall screen.
NOTE: You cannot turn the Archive option off
for asystole and ventricular fibrillation.

Alarm Limits Shortcut

Each parameter setup menu has an alarm menu selection which accesses the Alarm
Limits table, targeting associated parameters on the Alarm Limits table. Exiting the
Alarm Limits table returns you to the parameter setup screen.

ST and Arrhythmia Alarms

ST and arrhythmia parameters have their own alarm limits configuration screens,
which you can access by selecting the
alarm limits table (see page 5-6). Refer to chapters 10, and 9, for more information
about ST and arrhythmia tables.

Alarm History Table

The monitor stores up to 50 physiological alarm events for each patient. Events are
deleted when the patient is discharged. Data is stored in the monitor and remains with
the patient during Pick and Go transports. Data also survives power shutdowns. The
Alarm History table records all high-priority and medium-priority alarms, every
activation and deactivation of cardiac bypass mode, every change of patient category,
and records whenever you perform an
To access the Alarm History table
1. Press the
2. Click on
5-14
fixed key.
Fast Access
to display the Alarm History table.
Alarm History
D
/D
ELTA
ELTA
Available Settings
• Store
• Record
• Str./Rec.
• OFF
or
control button at the bottom of the
ST
ARR
or
All Alarms Off
Alarm Silence
XL/K
APPA
.
VF8