Hemodynamic Pods
User Interface
User Controls:
Displays:
Connections:
Degree of protection against ingress
of water:
Physical Attributes
Size (H x W x D):
Weight:
Electrical Specifications
Power source:
Protection against electric shock:
Mode of operation:
Defibrillation protection:
Environmental Requirements
Temperature range:
VF8
D
Fixed keys (C.O. Start, IBP Zero, Wedge)
HemoMed:No LCDs
Hemo2: two 4-character LCDs.
Hemo4: four 4-character LCDs.
MPod – Quad Hemo: No LCDs
Hemo2: 2 invasive pressures, C.O., 2 temperatures, single
cable connecting pod to monitor
Hemo4, MPod – Quad Hemo, and HemoMed:
4 invasive pressures, C.O., 2 temperatures, single cable
connecting pod to monitor
Hemo2, Hemo4, and HemoMed: IPX0 (per IEC 60529)
MPod – QuadHemo: IPX1 (protected against harmful
effects of dripping water) per IEC 60529
Hemo2, Hemo4, and HemoMed:140 x 205 x 60 mm
(5.5 x 8.1 x 2.3 in)
MPod – QuadHemo: 205 x 110 x 80 mm (8.1 x 4.3 x 3.2 in)
Hemo2 and HemoMed:0.7 kg (1.6 lb.)
Hemo4: 0.9 kg (1.9 lb.)
MPod – Quad Hemo: 0.48 kg (1.1 lb)
NOTE: Hemo2 weight includes one transducer adapter
block. Hemo4 weight includes two transducer adapter
blocks. Both weights exclude mounting clamp.
For the MPod – Quad Hemo the weight includes four
transducer cables but excludes the mounting clamp and
rod.
Powered directly from the monitor
Type CF
Continuous
Per IEC 60601-2-34
Operating:
Hemo2, Hemo4, HemoMed:10 to 40 °C (50 to 104 °F)
MPod – QuadHemo: 0°C to 45°C (32°F to 113°F)
Storage:
Hemo2, Hemo4, HemoMed: –20 to 50 °C (–4 to 122 °F)
MPod – QuadHemo: –40°C to +70°C (–40°F to +158°F)
/D
XL/K
ELTA
ELTA
M
ONITORING
APPA
A
CCESSORIES
B-25