SvO2/CCO Monitors

An SvO2/CCO monitor measures cardiac output and related hemodynamic
parameters. This Dräger monitor accepts data via an MIB interface from an Abbott Q2
(Q2+) SvO2/CCO monitor or from the Edwards/Baxter Vigilance and Vigilance II
SvO2/CCO monitors.
NOTE: The bedside monitor calculates the Body Surface Area (BSA) using the
patient height and weight that was entered in the "Admit Patient" menu (see page 4-2). It
uses this BSA value to calculate the Index parameters (for example, ICI, CCI, SVRI etc)
for the Baxter MIB device. Although the Baxter device can calculate BSA internally and
send ICI and CCI values to the monitor, the monitor will still calculate the index values
based on its BSA.
The following table lists the parameters available for display.
Label
Parameters
BT*
Blood Temperature
CCO
Continuous Cardiac
Output
CCI
CCO Index
ICO
ntermittent Cardiac
Output
ICI
ICO Index
SaO2*
Arterial Oxygen
Saturation
SvO2
Venous Oxygen
Saturation
SVR
Systemic Vascular
Resistance
SVRI
SVR Index
VO2
O
2
DO2
O
2
SV
Stroke Volume
SVI
SV Index
EDV
End Diastolic Volume
EDVI
EDV Index
ESV
End Systolic Volume
ESVI
ESV Index
EF
Ejection Fraction
The Abbott Oxymetrix 3 SvO2/CCO monitor also provides CCO and SvO2. See the documentation
accompanying that device for information about these parameters and their calculation.
VF8
D
Units
°C
L/min.
L/min./m
L/min.
L/min./m
%
%
dynes x sec x cm
dynes x sec x cm
Consumption
mL/min
Delivery
mL/min
mL
mL/m
mL
mL/m
mL
mL/m
%
/D
XL/K
ELTA
ELTA
SvO2/CCO Monitor from
Q2, Vigilance and
Vigilance II
Q2, Vigilance and
Vigilance II
2
Vigilance and Vigilance II
Q2, Vigilance and
Vigilance II
2
Vigilance and Vigilance II
Vigilance and Vigilance II
Q2, Vigilance and
Vigilance II
-5
Vigilance and Vigilance II
-5
2
x m
Vigilance and Vigilance II
Vigilance and Vigilance II
Vigilance and Vigilance II
Vigilance II
2
Vigilance II
Vigilance II
2
Vigilance II
Vigilance II
2
Vigilance II
Vigilance II
APPA
S
O2/CCO M
V
ONITORS
28-17

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity delta xlInfinity kappa

Table of Contents