Micro2+® Spo2 Status Messages - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

®
MicrO2+
SpO
Message
Possible Cause
SpO
* > #
The patient's SpO
2
SpO
* < #
the current upper or lower alarm
2
limits.
PLS* > #
The patient's pulse rate is
PLS* < #
outside the current upper or
lower alarm limits.
PLS* Out of Range
The pulse rate is outside the
(High)
measuring range of the monitor
SpO
% > UL
The
2
the upper limit.
SpO
* Transparent Too much light is reaching the
2
sensor's light detector. This
condition usually occurs
because the sensor is off the
finger or because too much
ambient light is detected.
SpO
*
An invalid value was detected for
2
Unrecognized
the sensor calibration resistor.
Sensor
SpO
* Light
Insufficient light is reaching the
2
Blocked
sensor's light detector.
NOTE: With detached or partially
detached disposable sensors,
the light emitters and detectors
may have become misaligned.
SpO
* Artifact
Persistent artifact
2
SpO
* Weak signal The pulse amplitude is too low
2
due to one of the following
conditions:
- Poor perfusion (shock).
- Low body temperature.
VF8
D
Status Messages
2
* is outside
2
SpO
% is greater than
2
/D
XL/K
ELTA
ELTA
M
O2+® S
ICR
TANDALONE
Suggested Action
Check the patient and treat, if necessary.
• Make sure the sensor is properly
attached to the patient's finger.
• Remove the light source.
• Cover the sensor with opaque material.
• Make sure no ambient light can reach
the detector.
• Contact DrägerService.
• Check for a defective or unapproved
sensor.
• Replace the sensor.
• Contact the hospital's technical
personnel or DrägerService.
• Make sure the light sensor is properly
attached to the patient's finger, and that
the finger is free of blocking substances.
• Check for a defective sensor and
replace, if necessary.
• Make sure the SpO
attached to the patient, the monitoring
site is free of patient motion, and all
cables are properly connected.
• Contact DrägerService.
• Check the patient.
• Make sure the SpO
attached to the patient and all cables are
properly connected.
• Relocate the sensor to another
extremity.
• Contact DrägerService.
APPA
P
O
ULSE
XIMETER
sensor is properly
2
sensor is properly
2
17-23