Dräger Infinity Delta Instructions For Use Manual page 162

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

8 ECG
H
AND
EART
Menu
Description
Selection
ECG Report
Allows you to set up and
Setup
request Rest ECG Analysis
reports on a laser printer
(provided the monitor is on the
network and is connected to a
Infinity CentralStation, has the
Aries option installed, and a 12-
lead MultiMed pod is
connected)
NOTE: To request a report, you
must first configure it at the
central station (see the Biomed
chapter in the Infinity
CentralStation Instructions for
Use for detailed information).
NOTES:
• When in neonatal mode, bradycardia is a low-priority heart rate alarm.
• A bradycardia alarm, which is a high-priority alarm, can be configured independently of the
low HR alarm, which is a medium-priority alarm.
Brady
Sets the bradycardia detection
Detection
limit.
Brady
Allows you to store and/or
Archive
record automatically a
bradycardia alarm event. You
can later review stored alarms
on the Event Recall screen.
8-22
R
ATE
The Brady Alarm submenu
(Only visible in neonatal mode)
This submenu allows you to configure the following functions.
D
/D
ELTA
ELTA
Available Settings
• Report – Generates Rest-ECG report
• Admit – Allows you to enter patient data (see
chapter 4, Admission, Transfer, and
Discharge.)
• Sex – Unknown, Male, Female
• Race – Unknown, Caucasian, Asian,
African origin, Other
• Medication 1 and Medication 2 – Unknown,
No Meds, Digitalis, Beta Blocker, Quinidine,
Procainamide, Amiodarone, Disopyramide,
Lidocaine, Other Antiarrhythmics, Diuretic,
Psychotropic, Steroid, Other Meds
• Condition 1 and Condition 2 – Unknown,
Normal, Infarction, Ischemia, Hypertension,
Congenital HD, Valvular HD, Pericarditis,
Respiratory Disease, Endocrine Disease,
Pacemaker, Pulm. Embolism, Post Op
Changes, Cardiomyopathy, Other
NOTE: (Available when Aries option is installed)
None, Chest Pain, Chest Pain Increase,
Chest Pain Gone, Chest Pain Decrease,
Routine ECG, Treatment Change, Start
Thrombolytics, Post Thrombolytics, ECG
Post Intervention, Post PTCA, Post CABG,
Position: Supine, Left Side, Right Side, Up
Right, Front; ECG Right Sided; V7-Vx ECG
• OFF
• 20 - 100 bpm in 5 bpm intervals.
• OFF
• Record
• Store (Default)
• Str./Rec.
XL/K
APPA
VF8

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity delta xlInfinity kappa