The Pick and Go patient transport system allows the monitor to travel with the patient
to different care stations within the hospital. By uploading setups from the IDS at the
new care station, the monitor adapts to its new clinical "home" (OR, ICU, CCU, etc.)
while retaining patient data.
The Pick and Go scenario below describes how monitored information follows the
patient from an ICU to the operating room.
1. Undock the monitor in the ICU — the monitor retains the ICU and patient
setup.
2. Transport the monitor with the patient— the monitor continues to use the
setups configured at the ICU bedside.
3. Dock the monitor in the OR — the monitor uploads the OR default monitor
settings.
4. Undock the monitor in the OR — the monitor retains the OR and patient
setups.
5. Return transport — the monitor continues to monitor the patient using the OR
monitoring configuration.
6. Redock the monitor at the ICU bedside — the monitor uploads the ICU
monitoring defaults at the bedside and resumes bedside monitoring. Patient
settings are not affected.
For more information on managing setups during Pick and Go operations, see page 2-
11. For information about setups management (including restoring patient or patient
and monitor settings), see page 2-9.
NOTE: A monitor may automatically change to wireless mode during Pick and Go, if
certain criteria are met (see page 3-10).
VF8
D
P
ICK AND
Transport
O
/D
XL/K
ELTA
ELTA
G
T
(D
O
RANSPORT
Delta/Delta XL
(
APPA
/D
XL
ELTA
ELTA
ONLY
only)
)
3-5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity delta xlInfinity kappa

Table of Contents