Monitor Without Masimo Or Nellcor Logo - Status Messages - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

17 P
O
ULSE
XIMETRY

Monitor Without Masimo or Nellcor Logo - Status Messages

Dräger OxiSure (via the MultiMed cables)
Message
SpO
> #
2
SpO
< #
2
PLS > #
PLS < #
SpO
< Low limit
2
value - 20%
PLS Out of Range
(Low/High)
SpO
Transparent
2
SpO
Unrecognized
2
Sensor
SpO
Light Blocked
2
SpO
Artifact
2
17-14
(S
O2)
P
Possible Cause
The patient's SpO
falls outside
2
the current upper or lower alarm
limits.
The patient's pulse rate falls
outside the current upper or
lower alarm limits.
The patient's SpO
falls below
2
the current lower alarm limit by
20% or more.
The pulse rate is outside the
measuring range of the monitor
Too much light is reaching the
sensor's light detector.
This condition usually occurs
because the sensor is off the
finger or because too much
ambient light is detected.
The monitor does not recognize
the sensor connected as valid.
Insufficient light is reaching the
sensor's light detector.
NOTE: With detached or
partially detached disposable
sensors, the light emitters and
detectors may have become
misaligned.
Persistent artifact
D
/D
ELTA
ELTA
Suggested Action
Check the patient and treat, if necessary.
• Make sure the sensor is properly
attached to the patient's finger.
• Remove the light source.
• Cover the sensor with opaque material.
• Make sure no ambient light can reach
the detector.
• Contact DrägerService.
• Check for a defective or an unapproved
sensor.
• Replace the sensor.
• Contact the hospital's technical
personnel or DrägerService.
• Make sure the sensor is properly
attached to the patient's finger, and the
finger is free of blocking substances.
• Check for a defective sensor and
replace it, if necessary.
• Make sure the SpO
attached to the patient, the patient is not
moving, and all cables are properly
connected.
• Contact DrägerService.
XL/K
APPA
sensor is properly
2
VF8

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity delta xlInfinity kappa

Table of Contents