Alarm Silence; Dismissing Alarm Messages - Dräger Savina 300 Instructions For Use Manual

Hide thumbs Also See for Savina 300:
Table of Contents

Advertisement

Alarms

Alarm silence

The acoustic alarm signal can be suppressed for a
maximum of 2 minutes.
If an alarm with a higher priority occurs during this
time, the alarm signal sounds once.
If the fault triggering the alarm is not eliminated
after 2 minutes, the alarm signal sounds again.
The alarm signal cannot be suppressed in the
following situations:
– During the alarm Standby mode activated
– During the device check when testing the alarm
signals

Dismissing alarm messages

After the fault has been eliminated, the alarm signal
stops. High-priority alarm messages continue to be
displayed and need to be dismissed.
 Press the
Alarm reset key.
110
Suppressing the alarm signal
 Press the
Audio paused 2 min. key.
Savina 300 displays the
remaining time for the suppressed alarm signal in
the header bar.
Reactivating the alarm signal
 Press the
Audio paused 2 min. key.
Instructions for use Savina 300 SW 5.n