Dräger WaterLock 2 Quick Start Manual

Dräger WaterLock 2 Quick Start Manual

Advertisement

Quick Links

QUICK START GUIDE
WaterLock® 2 /
Connecting the water trap
1.
Remove the new water trap from the packaging.
Write the current date on the front of the water trap.
2.
NOTE
The 2nd step is not necessary if the water trap Infinity ID WaterLock 2
is used together with Dräger devices with Infinity ID technology as the
maximum period of use of the water trap is automatically monitored.
3.
Hold the water trap
the holder or adapter until it audibly clicks into place.
4.
Connect the sample line
of the water trap.
Replacing the water trap
When the maximum period of use of 4 weeks is reached.
If there is an error message on the patient monitor or the anesthesia workstation.
If emptying does not rectify the error message of the patient monitor or the anesthesia workstation.
If it is severely soiled.
If the swelling seals in the connections turn blue.
To connect the new water trap, please follow the instructions from the chapter
"Connecting the water trap" above.
WaterLock® 2. 12 pcs.
Infinity ID WaterLock® 2, 12 pcs.
Sample line, disposable, 10 pcs.
Dräger. Technology for Life
Infinity ID WaterLock® 2
(A)
at the fluted grips and insert it into
(B)
to the Luer Lock connection
®
Please note: This Quick Reference Guide is not a replacement or substitute for the
instructions for use. Any use of the medical device requires full understanding and strict
observation of the instruction for use.
B
A
Order no.
6872130
6872020
8290286

Advertisement

loading

Summary of Contents for Dräger WaterLock 2

  • Page 1 Write the current date on the front of the water trap. NOTE The 2nd step is not necessary if the water trap Infinity ID WaterLock 2 is used together with Dräger devices with Infinity ID technology as the maximum period of use of the water trap is automatically monitored.
  • Page 2 Emptying the water trap Remove the sample line from the water trap. Press the grips of the water trap and pull outwards. Insert an empty syringe with a filling volume of at least 20 mL without canula into the blue connector. -20ml- Extract water, remove syringe and dispose of full syringe as infectious hospital waste.