Overview; Precautions - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

17 P
O
ULSE
XIMETRY

Overview

Dräger offers three algorithms for SpO
Masimo SET
Nellcor Oximax
If installing VF8 software on monitors that do not have a Masimo SET logo
(see page 17-4) or a Nellcor OxiMax logo (see page 17-5), Dräger OxiSure
(with connection via the MultiMed cables) is available.
These technologies enhance the quality of SpO
the percentage of functional hemoglobin saturated with oxygen (% SpO
patient's arterial blood. An SpO
infrared light. The monitor uses the difference between the two measurements to
calculate the percentage of saturated hemoglobin. Because light absorption varies with
blood volume and blood volume varies with pulse rate, the monitor can also derive a
pulse rate (PLS).
The SpO
sensor, available for adult, pediatric, and neonatal patients is connected to
2
the monitor via the various MultiMed and NeoMed cables, or the Infinity
SET SpO
SmartPod
2
NOTE: For a complete list of Dräger provided SpO
product, see page C-13.

Precautions

Refer to the "Safety Considerations" section starting on page 6, at the beginning of
these Instructions for Use, for general precautions regarding safe device operation.
SpO
measurements are particularly sensitive to the pulsations in the artery and the
2
arteriole. Measurements may not be accurate if the patient is experiencing shock,
hypothermia, anemia or has received certain medications that reduce the blood flow in
the arteries.
17-2
(S
O2)
P
monitoring:
2
®
®
sensor measures the absorption levels of red and
2
®
®
, or the Infinity
Nellcor Oximax
D
/D
ELTA
ELTA
monitoring, allowing you to measure
2
®
SpO
SmartPod
2
accessories available with this
2
XL/K
APPA
) in the
2
®
Masimo
®
pod.
VF8