9 A
RRHYTHMIA
Label
Event and Beat Classification
SVT
Supraventricular Tachycardia: N or more consecutive normal beats, with a beat-to-beat
rate greater than or equal to the SVT setting
CPT
Ventricular Couplet: Sequence of beats with the pattern: normal, PVC, PVC, normal
BGM
Ventricular bigeminy: Sequence of beats with the pattern: normal, PVC, normal, PVC,
normal
Sinus Tachycardia: N or more consecutive normal beats, with a beat-to-beat rate ≥ TACH
TACH
rate setting
Sinus bradycardia: 8 or more consecutive normal beats, with an average rate ≤ sinus
BRDY
bradycardia rate setting
NOTES:
• When in neonatal mode, bradycardia is a low heart rate alarm
• Brady Alarm (neonate only), which is a high alarm, can be configured independently of
the low HR alarm, which is a medium alarm.
Pause: Sequence of two beats classified as normal or PVC, with interval ≥ pause rate
PAUS
value in seconds (±100ms)
ARTF
Artifact: More than 50 % of beats in the last minute classified as questionable
1
Certain ventricular tachycardias have sinusoidal waveforms closely resembling those of
ventricular fibrillation. Because of the similarity of these waveforms, the monitor may classify
such types of ventricular tachycardia as ventricular fibrillation, the more serious of the two
conditions.
2
"N" is the event count set in the Arrhythmia setup table's count column.
3
A PVC or other abnormal beat breaks the analysis sequence and restarts analysis.

Automatic Learning and Relearning

After you connect ECG cables to the patient, the monitor begins to learn a reference
template whenever you execute any of the following tasks:
Turn on the monitor
Exit standby mode
Click on
Change the top-channel lead (ECG 1), or change the ECG2 channel lead
during ECG1&2 processing
Change a 12-lead configuration
9-4
M
ONITORING
2,3
2
ARR Monitoring
D
/D
ELTA
or
QRS/ARR Select
XL/K
ELTA
APPA
.
VF8

Hide quick links: