Status Messages - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Status Messages

Message
Check Printer
[Primary/Secondary]
Recorder not
connected
[Primary/Secondary]
Recorder not
assigned
[Primary/Secondary]
Recorder door open
Timed recording
started
Recording not
accepted
VF8
D
Possible Cause
The printer is not connected.
The recorder is disconnected or
the connection is poor.
You have not specified a
recorder.
The paper door is open.
The recorder is currently printing. Let the recorder finish printing.
The recorder does not
understand the print request.
/D
XL/K
ELTA
ELTA
Suggested Action
Check the printer connection.
• Connect a recorder and verify that it
is appropriately assigned.
• Inspect the cable, replace it, if
necessary.
Specify a recorder from the R50
Assign menu.
Close the door to the recorder paper
compartment.
Try again; call the hospital's technical
personnel.
APPA
S
M
TATUS
ESSAGES
7-13