Dräger Infinity Delta Instructions For Use Manual page 55

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

The Main Menu
Menu Item
Description
Parameters
Accesses the parameters' setup menus.
Alarm Limits Opens the alarm limits table
This submenu allows you to layout the main screen. To access the second page of this menu, click
on the down arrow at the bottom of the screen. Click on the up arrow to return to this page.
Parameter
Allows you to modify the order of
Priority
parameters displayed on the Main Screen
without regard to whether or not that
parameter is connected.
NOTES:
• Parameters must be assigned a priority;
whether they appear on the main screen
depends on their priority and on the
number of channels configured to
display waveforms (see Max. Channels
on page 2-6).
• In Automatic display mode, connected
parameters are displayed according to
their priority in the Parameter Priority
list. If all available channels are filled, a
higher priority parameter will NOT
"bump" lower priority parameter boxes
off the Main Screen when the
associated device is connected. In order
to display the parameter, the user must
double-click on the parameter in the
Parameter Priority list.
• In OR Mode, if all available channels are
filled, a higher priority parameter will
"bump" lower priority parameter boxes
off the Main Screen when the
associated device is connected.
VF8
D
The Monitor Setup submenu
Main Screen
/D
XL/K
ELTA
ELTA
C
ONFIGURING THE
Available Settings
• Click on one of the listed parameters
to access its setup menu: (for
example, ECG, ARR, ST, ART, PA
CVP, RA, SPO2, NBP, RESP,
etCO2, C.O.
NOTE: The parameters that appear in
this list depend upon the monitor's
configuration and which devices are
connected.
• Click on Exit to return to the first
column of the Main menu.
See page 5-6
To change a parameter's display order:
1) Scroll to Display Mode.
2) Select Manual.
3) Click on Parameter Priority to
highlight the first listed parameter.
NOTE: Parameters are numbered
according to their priority. The display
is color coded: A green parameter
label indicates that the associated
parameter device (for example, an
NBP hose/cuff) is connected to the
monitor. A white label indicates that the
device is not connected.
4) Scroll the list to the parameter you
wish to move and click.
5) Move the parameter to its new place
using the rotary knob.
6) Click to confirm the parameter's new
position on the list.
7) Click again to return to Parameter
Priority.
8) Scroll to the arrow at the upper left
corner of the menu to exit, or
continue to set up other Main
Screen submenu functions.
APPA
M
ONITOR
2-5