Infinity etCO
2
WA R N I N G : D o n o t a u t o c l a v e o r s t e r i l i z e t h i s d e v i c e .
CAUTION:
Do not spray or pour any liquid directly on the pod, accessories or
consumables.
Do not use caustic or abrasive cleaners, or harsh solvents, including
petroleum-based or acetone solutions, to clean the device.
Microstream ® etCO
not to be reprocessed. Do not attempt to clean, disinfect, sterilize or flush out
the FilterLine as this can damage the monitor
1. To clean the pod's surfaces, lightly dampen a cloth with a 70% alcohol
solution.
2.
Wipe all surfaces.
Alcohol wipes may also be used. The frequency of the cleaning procedure should be
determined by hospital policy.
FiO
2
Clean the FiO
sensor by wiping external surfaces with a cloth lightly dampened with
2
a mild detergent or isopropyl alcohol. Disinfect the external surface of the FiO
using a cloth moistened with ethanol.
CAUTION: Do not autoclave, gas sterilize, or irradiate the oxygen sensors. Do not
clean the sensor with chemicals other than alcohol or a mild cleaning agent.
VF8
D
Microstream SmartPod
consumables are designed for single patient use and are
2
/D
XL/K
ELTA
ELTA
APPA
F
O2
I
sensor
2
29-11