etN2O > 82%
MultiGas Sample Line
Occlusion
MultiGas Zero in
Progress
MultiGas Zero
Accepted
MultiGas Zero Failed
Check Watertrap/
Sample Line
MultiGas Too Warm
MultiGas Data Invalid
MultiGas Warming Up The Scio module is warming up and
MultiGas
Incompatible
MultiGas
Initialization
VF8
D
Condition
The module has not yet identified or
cannot identify the agent because
of one of the following conditions:
• Unknown agent (for example, not
HAL, DES, ISO, SEV or ENF)
• The agent concentration is too
low
• The vaporizer is leaking
• Traces of disinfectant are present
etN
O > 82 %
2
• The Scio module sample line is
occluded
• The water trap is full, defective, or
not installed.
The zeroing of the Scio module is in
progress.
The zeroing was successful.
• Zero performed using the wrong
gas
• Occlusion or leak present
• HW problem
The water trap is full or the sample
line is blocked.
• The fan port is blocked.
• Hardware problem.
Communication problem
is operating at reduced accuracy
Scio H/W or S/W incompatibility
The device is initializing.
/D
XL/K
ELTA
ELTA
M
AINTENANCE AND
Suggested Action
• (For Scio Four Oxi or Scio Four)
select agent manually.
• Check/replace vaporizer.
• Check N
O concentration in the
2
fresh gas flow.
• Flush.
• Check sample line and replace, if
necessary.
• Check water trap, replace or install,
if necessary.
Wait
Informational message; no action is
required.
• Verify that anesthetic gases have
not contaminated the surrounding
atmosphere.
• Check for leaks and occlusions.
• Call the manufacturer.
• Replace the sample line.
• Empty or the replace water trap
(see page 23-25).
• Clear/unblock the port.
• Call the manufacturer.
• Unplug and re-plug the Scio
module.
• Power-cycle the monitor or undock
and redock monitor.
• Call the manufacturer.
Do not rely on the Scio module
values.
• Check the version numbers.
• Call the manufacturer.
Wait