Dräger Infinity Delta Instructions For Use Manual page 271

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Message
C.O. No
Temperature
Change
C.O. Use Cooler
Injectate
C.O. Injectate Set to
<IT value>!
C.O. Average Saved The C.O. average has been saved.
C.O. Transducer
Unplugged
C.O. Poor Baseline
C.O. Pod Fault -
Bad Ref.
C.O. Catheter Fault
- Bad Ref.
C.O. Check
Injectate Probe
VF8
D
Possible Cause
The detected temperature change was
< 0.1 °C, because of the following
conditions:
• C.O. START was pressed but no
injection was made
• The injectate volume was too small
• Defective catheter
• The injectate temperature was too
warm.
• < 5 °C difference exists between
patient's blood temperature and the
injectate temperature.
• Injectate temperature of greater than
30°C.
C.O. START was pressed, but no
injectate probe is connected.
A cable or transducer has become
disconnected.
The temperature curve did not return
to baseline within 30 seconds of
pressing C.O. START because of the
following conditions:
• Unstable patient temperature
• Defective catheter, cable, or
cartridge
The pod reference resistance is either
too high or too low.
• The catheter reference resistance is
too low.
• Unknown catheter type.
the injectate temperature probe is not
connected or was disconnected during
a measurement.
/D
XL/K
ELTA
ELTA
Suggested Action
• Repeat the measurement.
• Use a larger injectate volume.
• Repeat the measurement. If
problem persists, replace the
catheter.
• Use colder injectate.
Use a cooler injectate.
Connect injectate probe.
None required.
Reconnect the cable or transducer.
If message persists, replace
defective part.
• Follow hospital procedures.
• Repeat the measurement. If
message persists, replace the
defective components.
Remove and reconnect the pod.
Repeat the measurement. If
message persists, replace the pod
and contact DrägerService.
• Check the catheter and replace if
defective.
• Contact the hospital's technical
personnel or DrägerService.
Connect the probe and repeat the
measurement.
APPA
S
M
TATUS
ESSAGES
15-15

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity delta xlInfinity kappa