Biomed functions are described in the following table:
Quick Reference – Biomed Menu
Menu Item
This menu displays clinical and technical diagnostic records. Open the Biomed menu (page 2-21)
and click on Logs, then follow the procedures described in this table.
Component Log
Status Log
Diagnostic Log
CPS/IDS Diag. Log
FE Diag. Log
Copy All Logs
Print Log
and intended for the hospital's technical personnel or DrägerService personnel only.
Test Pulse
2-22
Description
Logs
The
The monitor maintains read-only
logs for principal components or
devices. These logs include a part
number, revision numbers, serial
numbers, software version, and
compatibility information.
Displays information about your
current software and hardware
versions.
Captures data about hardware and
software performance relating to the
monitor's operation.
Captures data about CPS and IDS
performance relating to the monitor's
operation.
Captures and displays data about
front-end performance related to the
monitor's operation.
Downloads status logs and
diagnostic logs to a memory card.
Prints an expanded version of the
Diagnostic Log to an Infinity network
laser printer
Service
The
The Service menu is password-protected
Test Pulse
Tests the ECG and EEG signal
clarity and display.
D
/D
ELTA
ELTA
Settings/Procedures
Submenu
1) Scroll to the component you
wish to inspect and click on it.
The log appears with the exit
arrow already selected.
2) Return to the Logs submenu by
clicking again.
The display is read-only.
Display is read-only.
• Click on the down arrow at the
bottom of screen to scroll
through the display.
Display is read-only.
• Click on the down arrow at the
bottom of screen to scroll
through the display.
Display is read-only.
• Click on the down arrow at the
bottom of screen to scroll
through the display.
Click on Copy All Logs. A
confirmation message appears in
the message area to indicate the
download is complete.
Click on Print Log.
submenu
Clicking on Test Pulse causes the
following to happen:
• Injects a 300 ms pulse into the
ECG waveform (1 mV on leads I
and III, 2 mV on lead II).
• Superimposes a 100 µV pulse on
all EEG channels for 200 ms.
XL/K
APPA
VF8