14 I
B
NVASIVE
Duplicate Label Conflicts Between Pods
Conflicts also occur when two Hemo2, Hemo4, MPod – Quad Hemo or PiCCO pods
are configured for the same labels and transducers are connected to the pods.
Scenarios for duplicate label conflicts may vary. They may be detected when a second
pod is connected to an operating monitor, in which case the label of the first connected
pod prevails. If two pods storing duplicate labels are connected to a monitor prior to
start-up, the pod connected to the lowest-numbered pod port on the monitor (Aux/
Hemo2 or Aux/Hemo3) has priority, while the other pod is assigned an automatic
label (P1a
P3d).

Pulmonary Wedge Pressure Display

NOTE: PWP is not available with the PiCCO pod.
WA R N I N G : D u r i n g w e d g e p r e s s u r e m e a s u r e m e n ts :
The monitor has a special display that accommodates taking pulmonary wedge
pressure measurements. The monitor averages the PA waveform values for 10
seconds and calculates a wedge pressure value (PWP). During the measurement, the
PA parameter box shows no systolic or diastolic values, and PA alarms are disabled.
14-18
P
LOOD
RESSURE
A l a r m m o n i t o r i n g f o r i n v a s i v e p r e s s u r e s , i f e na b l e d , i s
t e m p o r a r i l y d i s a b l e d t o p re v e n t n u i s a n c e a l a rm s . T he
pa r a m e t e r b o x d oe s n o t d i s p l a y a c r o s s e d b e l l i c o n
b e c a u s e a l a r m m o n i t o r i n g i s a u t o m a t i c a l l y e n a b l e d
u po n c o m p l e t i o n o f a w e d g e p r e s s u r e m e a s u r e m e n t .
F o r t h e s a f e t y o f t he pa t i e nt k e e p t h e b a l l o o n - i n f l a t i o n
t i m e t o t h e m i n i m u m n e c e s s a r y t o a c q u i r e a n a c c u r a t e
P W P v a l u e . P r o l o n g e d i n f l a t i o n o f t h e b a l l o o n c a n
r e s u l t i n p u l m o n a r y h e m o rr h a g e o r i n f a rc t i o n .
D o n o t o v e r - i n f l a t e t h e b a l l o o n b e c a u s e a n o v e r - i n f l a t e d
b a l l o o n c a n r up t ur e t h e pu l mo na ry a rt e r y.
T he PA c a t h e t e r m a y m o v e i n t o t h e w e d g e p o s i t i o n
b e f o r e t h e b a l l o o n i s i n f l a t e d. O n e s i g n o f t h i s "c a t h e -
t e r d r i ft " i s t h a t t h e P W P w a v e f o rm b e c o m e s w e d ge
s h a p e d . F o l l o w y o u r h o s p i ta l 's c l i n i c a l g u i d e l i n e s t o
c o r r e c t t h e c a t h e t e r 's p o s i t i o n.
D