Dräger Infinity Delta Instructions For Use Manual page 574

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

B T
D
ECHNICAL
Acetone
Methane
End-Tidal CO
Parameter display:
Measurement method:
Measurement modes:
Measurement range:
Accuracy
(full accuracy from the time the first
value appears on the screen):
Drift of measurement accuracy:
Quantitative effects of humidity and
condensate on accuracy
Adverse effects on performance of
cyclical pressures up to 10 kPa (100
cmH2O)
Effect of barometric pressure on
accuracy
Sampling flow rate:
Gas leakage:
Alarm limit resolution:
RRc range (pod):
RRc accuracy
Rise time (10 to 90%):
Delay time:
Total system response time:
Compensation:
Calibration interval:
B-48
ATA
0.1
1.0
(etCO
)
via I
NFINITY
2
2
etCO
Dual wavelength, non-dispersive infrared absorption
Adult, Pediatric and Neonatal: Sidestream only
0-99 mmHg CO2 partial pressure
0 to 38 mmHg:±2 mmHg
39 to 99 mmHg:±5 % of reading +0.08 % for every
≤2.0 mmHg over 24 hours
Full accuracy specifications are maintained for all non-
condensing humidity levels.
No effect on accuracy
No effect on accuracy
50 +15 mL/min. or –7.5 mL/min.
Not applicable
etCO
RRc: 1 breath/min
0-150 breaths/min
0 to 70 bpm:±1 bpm
71 to 120 bpm:±2 bpm
121 to 150 bpm:±3 bpm
≤190 ms
2.7 sec. maximum @ flow 50 mL/min with standard
FilterLine™
Typical 4 s
BTPS (standard correction used by Microstream
capnography during all measurement procedures for body
temperature, pressure, and saturation
Initially calibrate after 1,200 operating hours, then once a
year or after 4,000 operating hours, whichever comes first.
The initial calibration should not occur before 720 hours of
use
D
/D
ELTA
etCO
Microstream
2
, iCO
, Respiration Rate (RRc)
2
2
1 mmHg (above 38 mmHg)
, iCO2: 0.1 mmHg (0.01 kPa)
2
XL/K
ELTA
APPA
pod
VF8

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity delta xlInfinity kappa

Table of Contents