Status Messages - Dräger Infinity Delta Instructions For Use Manual

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

14 I
B
NVASIVE

Status Messages

Message
Possible Cause
<xx> S <#>
The pressure value is outside the
alarm limits.
<xx> D <#>
<xx> M <#>
<xx> Out of
The pressure signal is out of the
Range (High)
measurement range.
<xx> Out of
Range (Low)
<xx> Please
The IBP zero value stored in the
Check Zero
monitor may not correspond to
the peripheral device.
<xx> Static
Static pressure was detected on a
Pressure
pulsatile signal, owing to one of
the following conditions:
• A physiological condition, for
• The transducer was turned off to
• A catheter tip lodged against a
• A clot on the catheter tip.
<xx>
The pressure transducer for the
Unplugged
specified parameter is either
unplugged or defective.
<xx> Zero
The pressure transducer for the
Required
specified parameter requires
zeroing.
<xx> Zero
The zeroing of the transducer was
Accepted
successful
<xx> Did Not
The zeroing of the transducer
Zero
failed because of one of the
following conditions:
• Excessive signal noise
• A non-static waveform
<xx> Did Not
The zeroing of the transducer
Zero - Offset
failed because static pressure
Error
was too high or too low.
NOTE: <xx> represents the IBP parameter label associated with the displayed message.
14-20
P
LOOD
RESSURE
example, asystole.
the patient.
vessel wall.
D
/D
ELTA
Suggested Action
• Check the patient and treat, if necessary.
• Access the Alarm Limits menu and change
the alarm limits.
• Check the equipment and replace it, if
necessary.
• Check the patient and treat, if necessary.
• Access the Pressure Labels menu and
assign the correct label.
• Check the equipment and replace it, if
necessary.
Zero the transducer.
• Check the patient and treat,if necessary.
• Open the system to the patient by turning the
stopcock.
• Follow hospital procedures for dislodging
catheters.
• Follow hospital procedures for clotted
catheters.
• Active pressure: reconnect or replace the
cable.
• Inactive pressure: turn off alarms.
Zero the transducer.
No action is required.
• Keep all tubing motionless, then rezero.
• Change the transducer.
• Check the stopcock, then rezero.
• Rezero the transducer.
• Loosen and retighten the transducer dome,
then rezero the transducer.
• Replace the transducer.