Low perfusion
(6)
performance:
(6)The Masimo SET® pod has been validated for low perfusion accuracy in bench top testing
against a Biotek Index 2 simulator and Masimo's simulator with signal strengths of greater than
0.02% and a % transmission of greater than 5% for saturations ranging from 70 to 100 %. This
variation equals plus or minus one standard deviation.
Pulse Oximetry
Parameter display:
Measurement range:
Measurement accuracy, Adult mode (1, 2):
Saturation (%SpO2):
0 to 60% not specified
60 to 80% sensor-specific as follows:
Nellcor :
OxiMAX MAX-A, OxiMAX MAX-AL, OxiMAX MAX-P,
OxiMAX MAX-N, OxiMAX MAX-I, OxiMAX MAX-FAST
Nellcor :
SC-A, MAX-R, OxiCliq A, OxiCliq P, OxiCliq N, OxiCliq I,
D-YS, DS-100A, OXI-A/N, OXI-P/I
70 to 100% sensor-specific as follows:
Nellcor :
OxiMAX MAX-A, OxiMAX MAX-AL, OxiMAX MAX-P,
OxiMAX MAX-N, OxiMAX MAX-I, OxiMAX MAX-FAST
SC-A
Nellcor :
OxiCliq A, OxiCliq P, OxiCliq N, OxiCliq I
Nellcor :
D-YS, DS-100A, OXI-A/N, OXI-P/I
Nellcor :
D-YS with D-YSE Ear Clip, D-YS with D-YSPD Spot Clip
80 to 100% sensor-specific as follows:
Nellcor :
MAX-R
Pulse Rate (3) :
NOTES:
1) Since pulse oximeter measurements are statistically distributed, only about two-thirds of those
measurements can be expected to fall within ±1 A
2) The Infinity® Nellcor Oximax® SpO2 SmartPod™ pulse oximeter with Adult sensors has been
validated in human blood studies on healthy adult volunteers in induced hypoxia studies in the
range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation
equals ±1 A
rms
3) The pulse rate accuracy has been validated on healthy adult volunteers during induced hypoxia
studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This
variation equals ±1 A
B-44
ATA
Pulse amplitude:>0.02 %
Saturation (%SpO
and %transmission >5 %
Pulse rate:±3
(Nellcor Oximax SpO
Saturation (%SpO
%SpO
: 1 - 100%
2
Pulse rate:25 - 250 /min
25 - 250 bpm ±3 beats/min or ±3% (whichever is greater)
of the value measured by a co-oximeter.
of the pulse rate value measured by the ECG monitor.
rms
D
/D
ELTA
):±2
2
via Pod )
2
), pulse rate
2
of the value measured by a co-oximeter.
rms
XL/K
ELTA
APPA
±3
not specified
±2
±2.5
±3
±3.5
±3.5
VF8