Networked Alarms

The monitor can broadcast networked alarm messages to any compatible monitor or
central station within the network (typically in less than two seconds). However, a
central station will not annunciate or display any alarms for a connected bedside
monitor whose parameter alarms are turned off. In the Infinity network, you can also
group monitors into separate alarm groups to limit the number of messages posted at a
given device (see page 3-14).
Monitors connected to the network automatically relay alarms to the central station. If
the central station cannot acknowledge an alarm within 10 seconds (because of a
network interruption, for instance), the monitor displays the message, Network Alarm
Error
and sounds a tone at maximum volume (100 %). The alarm volume remains set
,
at 100% until you change it on the Alarm Limits menu (see page 5-6).
NOTE: The network distinguishes between unwanted network interruptions (offline
errors) and the deliberate undocking of a Delta/Delta XL bedside device during Pick and
Go transport operations. Removing a monitor from its IDS or Docking Station does not
trigger a network error alarm.

Alarm Priorities

The monitor has three alarm priorities: high, medium and low. Previously, Dräger has
referred to these as life-threatening (L-T), serious (SER), and advisory (ADV) alarm
conditions. You can define alarm priorities for arrhythmia and ST parameters only,
using the arrhythmia setup table (see page 9-6) or the ST alarms table (see page 10-
10). Each alarm priority has its own distinctive auditory alarm signal (alarm tone) and
color.
Alarm Priority
High
Medium
Low
VF8
D
Examples
Asystole, ventricular fibrillation
Neonatal apnea, patient safety conditions during
NBP measurements
Technical conditions such as misapplied
transducer or disconnected lead
/D
XL/K
ELTA
ELTA
APPA
A
P
LARM
RIORITIES
Alarm Color
Red
Yellow
White