2 M
S
ONITOR
ETUP
Setups Management
Menu function/item Description
Follow these procedures to save setups you have configured and named.
Save Setup
CAUTION:
Save
setups with care;
Saving a setup
overwrites an
existing setup.
Follow these procedures to restore setups you have configured, named, and saved.
Restore Setup
CAUTION: Restore
setups with care;
Restoring a setup
overwrites an
existing setup.
2-10
Saving Setups
NOTES:
• A Save Setup function is
also available on the
Independent Surgical
Display (see page 28-20).
Independent Surgical
Display configuration does
not affect monitor setups.
• You can save a setup on a
monitor that is not
connected to the network
only if you have defined the
setup as Default on the
Unit Manager menu under
Rename Setups. All other
setup options are ghosted.
Restoring Setups
NOTES:
• You can access Restore
Setups quickly as follows:
1) Press the Menu fixed
key.
2) Click on Restore
Setups.
3) Follow Steps 4-6 at
right.
• Save Setup and Restore
Setup menu selections are
also available on the
Independent Surgical
Display (see page 28-20).
The configuration of the
Independent Surgical
Display (ISD) Save or
Restore menu does not
affect monitor setups.
• If a monitor is not
connected to a network,
you cannot restore user-
defined setups.
• You can also restore
setups from the Main
menu.
D
/D
ELTA
ELTA
Reference/Procedures
1) Open the Unit Manager menu (see page
2-9).
2) Click on Save/Restore.
3) Click on Save Setup.
4) Click on the name of the setup you wish to
save. The monitor saves the setup with its
new configuration. A tone indicates the
monitor has successfully saved the setup.
1) Open the Unit Manager menu (see page
2-9)
2) Click on Save/Restore.
3) Click on Restore Setup.
4) Click on the name of the setup you wish to
restore.
5) Select the type of setting you wish to
restore:
• Monitor Settings Only to restore settings
configured on the Monitor Setup menu, or
• Patient and Monitor Settings to restore
monitor settings and settings configured on
the Patient Setup menu.
6) The monitor restores the setup and
returns to the Restore Setups menu; or
Indicates which parameters will be
removed to make room for restored setup.
7) Click on New Setup to remove the
indicated parameter(s) and restore the
selected setup, or Cancel to return to the
main screen.