Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Parameter
Predefined Alarm Range
PLS/PLS*
30 to 300 beats per minute
PVC/min
Adult and pediatric: 1 to 50 PVC per
minute (upper high limit only)
RA
-5 to 300 mmHg (-0.7 to 40.0 kPa)
RESP
Adult: 5 to 100 breaths per minute
Pediatric and Neonatal: 5 to 145
breaths per minute
RRc
5 to 145 breaths per minute
RRc*
5 to 90 breaths per minute
RV S/M/D
-5 to 300 mmHg (-0.7 to 40.0 kPa)
SpO
/SpO
2*
20 to 100%
2
2
NOTE: High and low SpO
(saturation) alarm conditions are
graded "Medium".
SpO
%
1 to 100% (high limit only)
2
ST Alarms
See "ST Alarms Table" on page 10-10.
VF8
D
2
/D
XL/K
ELTA
ELTA
A
S
(A
LARM
ETUP
Default
Default Alarm
State
Setting / Alarm Priority
Off
Adult:
Low
High
120 bpm
Pediatric:
Low
High
150 bpm
Neonatal:
Low
High
180 bpm
On
High: 10 PVC per minute
Off
Low
2 mmHg (0.3 kPa)
High
12 mmHg (1.6 kPa)
Off
Adult:
Low
5 bpm
High 30 bpm
Pediatric/Neonatal:
Low 20 bpm
High 80 bpm
Adult:
Adult:
Off
Low
5 bpm
Ped/
High 30 bpm
Neo:
Pediatric/Neonatal:
On
Low 20 bpm
High 60 bpm
Off
Low
5 bpm
High 30 bpm
Off
S:
Low 10 mmHg (1.3 kPa)
High 35 mmHg (4.7 kPa)
M:
Low
7 mmHg (0.9 kPa)
High 17 mmHg (2.3 kPa)
D:
Low
2 mmHg (0.3 kPa)
High 13 mmHg (1.7 kPa)
Adult:
Adult/Pediatric:
Off
Low 90% High 100%
Ped/
Neonatal:
Neo:
Low 85% High 95%
On
Off
Adult/Pediatric:
High 20%
Neonatal:
High 10%
APPA
L
T
LARM
IMITS
ABLE
45 bpm
50 bpm
80 bpm
5-11
)