Dräger Infinity Delta Instructions For Use Manual page 243

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Transducer cables plug into adapter blocks on the back of the pod and can be mounted
on the front panel. Slide the transducer into the slot nearest the associated pod
connector. With the MPod – QuadHemo, the transducers are attached to intermediate
cables which are permanently fastened to the back of the MPod.
1
2
3
To connect the hemodynamic pod to the monitor
1. Plug one end of the cable into the appropriate monitor connector.
Delta/Delta XL
1
Hemo2, Hemo4 or MPod – Quad Hemo connector (second
connector optional on Delta) or Picco pod connector
2
HemoMed connector
2. Connect the other end of the cable into the hemodynamic pod connection port.
VF8
D
Hemo4 Pod
Rear View
Transducer
Transducer cable
Adapter blocks
/D
XL/K
ELTA
ELTA
Kappa
APPA
H
S
ARDWARE
ETUP
Side View
14-9