Dräger Infinity Delta Instructions For Use Manual page 265

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Catheter
Injectate
Size
Volume
5F
5 cc
5F
3 cc
Injectate
Catheter Size
Volume
7.5F
10 cc
7.5F
5 cc
7.5F
3 cc
7F
10 cc
7F
5 cc
7F
3 cc
5F
5 cc
5F
3 cc
If you choose
Other
order to display or select the catheter size and injectate volume. Consult the
documentation included with the catheter for computation constants, and select one
that corresponds to the injectate volume and temperature being used.
VF8
D
BD/Ohmeda
Injectate Temperature (IT)
Sensor connected
IT = -5 °C to +16 °C
0.291
0.170
Arrow
Injectate Temperature (IT) Sensor connected
IT = -1 °C ( ±1 °C)
0.532
0.249
0.131
0.541
0.250
0.134
0.267
0.157
as the catheter type, you must enter a computation constant in
/D
XL/K
ELTA
ELTA
C.O. S
IT = 16 °C to 27 °C
0.316
0.188
IT = 24 °C ( ±1 °C)
0.586
0.265
0.155
0.601
0.273
0.156
0.303
0.192
APPA
- H
ETUP
ARDWARE
IT Sensor
disconnected
IT = 0 °C
0.279
0.160
15-9