If a wireless monitor loses contact with all access points and wireless transmission is
interrupted (for example, you remove the wireless card or the monitor is out of range),
the network generates an offline message and the monitor operates as a standalone
device.
If a wireless monitor regains contact with any access point (for example, you insert the
wireless card or the monitor is brought back into range) the normal monitoring state
will be restored and the offline message cleared within 40 seconds.
CAUTION: When the monitor is in wireless mode, patient data is not displayed at the
Infinity CentralStation while software upgrades, saved setups and card data transfers
are occurring.
Wireless During Pick and Go
A monitor in IDS mode will automatically switch to wireless mode when it is
undocked from the Docking Station, if:
a wireless card is inserted (see page 3-8).
and
all wireless settings are correct (contact your hospital's technical personnel).
When the monitor is re-docked to the Docking Station, it will automatically return to
IDS mode.
NOTE: If the setting Keep bed label is set to yes, the IDS bed label is maintained
when the monitor is undocked.
VF8
D
IDS
Bedside
Monitor
Wireless
Monitor
MVWS = Infinity CentralStation
IDS = Infinity Docking Station
AP = Access Point
/D
XL/K
ELTA
ELTA
MVWS
AP
IDS
Bedside
Monitor
APPA
W
N
IRELESS
ETWORK
3-11