Dräger Infinity Delta Instructions For Use Manual page 572

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

B T
D
ECHNICAL
Response time:
Site timer:
Temperature setting:
Sensor calibration:
Barometric pressure:
Heating power:
End-Tidal CO
Parameter display:
Measurement method:
Measurement modes:
Warm up:
Measurement range:
Accuracy:
Drift of measurement
accuracy:
Quantitative effects of
humidity and condensate on
accuracy
Adverse effects on
performance of cyclical
pressures up to 10 kPa (100
cmH2O)
Effect of barometric
pressure on accuracy
Calibration:
Compensation:
Sampling flow rate:
Gas leakage:
B-46
ATA
10 to 90 % response to a step change of O2: 20 s
10 to 90 % response to a step change of CO2: 20 s
0 to 8 hours or OFF (selectable)
37 to 45 °C, selectable
One-point calibration with CAL1 gas (fixed gas concentration of
20.9 % O2 and 5.0 % CO2)
500 to 800 mmHg, selectable
Measurement range:0 to 660 mW
Resolution:0.1 °C
Accuracy:±10 %
(etCO
)
via etCO
2
2
etCO2, iCO2, Respiration Rate (RRc)
Dual wavelength, non-dispersive infrared absorption
Adult and Pediatric:Mainstream and Sidestream
Neonatal:Mainstream only
≤5 min (at 25 °C)
0 to 99 mmHg CO2 partial pressure
0 to 40 mmHg:±2 mmHg
41 to 70 mmHg:±5 % of reading
71 to 99 mmHg:±8 % of reading
Stable over 24 hours, over full range of readings at atmospheric
pressure
≤1.6 mmHg over 8 hours
Full accuracy specifications are maintained for all non-condensing
humidity levels.
No effect on accuracy
No effect on accuracy
Verify once a day
Calibrate when moving the sensor from one module to another
Calibration time:< 20 s
Balance: User-selectable
Atm. pressure:Automatic or user-selectable
(540 to 800 mmHg)
180 ±12 mL/min (sidestream measurement mode)
<1.6 ml/min
D
/D
ELTA
module or pod
2
XL/K
ELTA
APPA
VF8