Appendix 2: Vitalert 2000 Alarm Messages - Dräger VITALERT 2000 Operator's Instruction Manual

Vital signs monitoring system
Table of Contents

Advertisement

APPENDIX 2: VITALERT 2000 ALARM MESSAGES

CLASS
ALARM
MESSAGE
W
HRT RATE LOW
C
HRT RATE HI
C
ART SYS LOW
C
CVP MEAN LOW
C
PA DIAS LOW
C
P2 SYS LOW
C
ART SYS HI
C
CVP MEAN HI
C
PA DIAS HIGH
C
P2 SYS HIGH
C
ECG PWR FAIL
A
ECG AC FAIL
A
ECG LL DISC
A
ECG RL DISC
A
ECG LA DISC
32
RETURN TO CD-ROM TABLE OF CONTENTS
CONDITION
Heart rate less than low
limit
Heart rate greater than
high limit
Arterial systolic less than
low limit
CVP mean less than low
limit
PA diastolic less than low
limit
P2 systolic less than
lowlimit
Arterial systolic greater
than high limit
CVP mean greater than
high limit
PA diastolic greater than
high limit
P2 Systolic greater than
high limit
No AC power and battery
low
AC power failure/
disconnect
ECG left leg lead
disconnect
ECG right leg lead
disconnect
ECG left arm lead
disconnect
AUDIO
ALARM
CODE
CONT
05H
INT
21H
INT
5FH
INT
7DH
INT
7FH
INT
87H
INT
69H
INT
80H
INT
82H
INT
88H
INT
9AH
SINGLE
90H
TONE
NONE
4FH
NONE
8CH
NONE
50H
ALARM
PRIORITY
31
21
21
21
21
20
20
20
20
20
11
7
2
2
2

Advertisement