Dräger Infinity Delta Instructions For Use Manual page 65

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

The Unit Manager Menu
Menu Item
Description
Extend All Alarms
Determines whether clinical personnel can
OFF
use the All Alarms Off fixed key to extend
the All Alarms Off time.
NOTE: This item only appears if your
monitor is configured for this feature
(contact your hospital's technical
personnel for more information).
Alarm
Validates the alarm conditions to limit
Validation
nuisance alarms due to artifact or motion/
movement by delaying time to alarm.
NOTE: When the Alarm Validation
setting is set to ON, the time to alarm from
the onset of a limit violation equals the time
for detection plus the designated alarm
validation signal delay. For HR, this time
may exceed the maximum of 10 seconds
as required by AAMI EC13 and IEC
60601-2-27.
Alarm validation times are:
HR
RESP
IBP
SpO
PLS
SpO
PLS*
All other parametersimmediately
1
SpO2 Alarm
Validates an SpO
Delay
requiring that the violation persists for
10 seconds (lower limit) before sounding
an alarm.
NOTE: Alarm validation must be turned
ON. If Alarm Validation is turned OFF,
SpO2 Alarm Delay is automatically
turned OFF and menu selection is
ghosted.
ASY/VF Alarms
Allows user to prevent disabling of ASY
and VFIB alarms.
WA R N I N G :
i s s e t t o O FF a nd A R R
m on it o r i ng i s s e t t o " OF F" , th e
m on ito r can no t ge ner ate ASY/ VF
a la rms.
VF8
D
Upper limitLower limit
6 sec
6 sec
14 sec
14 sec
4 sec
10 sec
6 sec
10 sec
2
6 sec.
10 sec
*
4 sec
4 sec
2
4 sec.
immediately
See SpO2 Alarm Delay below.
alarm condition by
2
W he n the HR alar m
/D
XL/K
ELTA
ELTA
Available Settings
• Enabled
• Disabled (default)
• ON – Enables the alarm
validation.
• OFF (default) – Disables the
alarm validation.
1
1
• ON – SpO
alarm conditions are
annunciated after it persists for
a period of 10 seconds.
• OFF – SpO
alarm condition is not validated
before annunciation.
• Always ON – ASY and VFIB
alarms are always active.
• Follow HR Alarms (default) –
ASY and VFIB alarms follow the
HR alarm settings.
APPA
S
M
PECIALTY
ENUS
or PLS lower limit
2
or PLS lower limit
2
2-15