Dräger Infinity Delta Instructions For Use Manual page 516

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

29 C
LEANING AND
Airway Adapter
CAUTION:
1. Rinse airway adapters in a warm soapy solution, then soak them in a liquid
disinfectant.
2. Dry with a lint-free cloth, making sure adapter windows are dry and free of
any residue before they are used.
Sidestream Sampling Pump (etCO
The etCO
module and pod contain a small pump that draws air from the nasal cannula,
2
through the sidestream airway adapter, and out the exhaust port. Suggested cleaning proce-
dures are outlined below.
WA R N I N G : U s e d s a m p l i n g t u b i n g , T- c o n n e c t o r s , a n d w a t e r
t r a ps c o u l d b e c on ta m i n a t e d a n d m u s t b e h a n d l e d a nd
d i s p o s e d o f w i t h c a r e . I n f e c t i o n h a z a r d m a y b e p re s e n t .
D i s p o s e o f t he s e i t e m s i n a c c o r d a n c e w i t h l o c a l r e g u l a t i o n s .
The following fluids are acceptable for cleaning:
Isopropyl alcohol.
A 5.25 % water solution (by weight) of sodium hypochlorite (bleach).
The following items are required for cleaning the etCO
A 60 cc catheter-tip syringe.
A 2-foot section of 1/8- or 3/16-inch tubing to drain off fluid after it passes
through the etCO
A receptacle to receive the fluid after it drains.
29-8
D
ISINFECTING
To reduce the risk of infection, remember that airway and flow sensors are for
single patient use only and cannot be sterilized. To avoid blocking the sampling
tube, do not attempt to sterilize the pod by immersing it in liquids.
Do not attempt to sterilize the etCO
it in liquids.
pump.
2
D
/D
ELTA
/Respiratory Mechanics pod by immersing
2
only)
2
2
XL/K
ELTA
APPA
module or pod:
VF8