Dräger Infinity Delta Instructions For Use Manual page 255

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Message
Possible Cause
<xx>
Mercury calibration is in progress. Complete the calibration before you begin
Calibrating
<xx> Cal.
Mercury calibration succeeded or
Accepted
the user-entered calibration factor
was accepted.
<xx> Cal.
The mercury calibration failed
Failed - Not
because the input pressure was
Static
not static.
<xx> Cal.
The mercury calibration failed
Failed - Out of
because the measured value was
Range
too high or too low.
<xx> Zero
During calibration, more than 5
before Cal.
minutes have elapsed since the
last successful zeroing of the
transducer.
<xx> H/W
IBP channel hardware failure.
Failure
Hemo Pod [n]
The hemo pod [1, 2, or 3] is not
Disconnected
connected to the monitor.
NOTE: <xx> represents the IBP parameter label associated with the displayed message.
VF8
D
/D
XL/K
ELTA
ELTA
Suggested Action
monitoring the patient.
No action is required.
• Make sure the transducer is closed to
patient.
• Check for leaks.
• Keep all tubing motionless.
• Rezero the transducer.
• Refer to the calibration procedures (see
page 14-5).
• Loosen and retighten the transducer dome,
then rezero the transducer.
• Replace the transducer.
• Make sure the transducer is zeroed, then
retry. If the retry fails, replace the transducer.
• If calibration requires a factor outside this
range, replace the transducer.
Zero the transducer.
• Check the hardware and replace it, if
necessary.
• Call the hospital's technical personnel or
DrägerService.
Check the cables and connections and
replace it, if necessary.
APPA
S
M
TATUS
ESSAGES
14-21