Dräger Infinity Delta Instructions For Use Manual page 467

Infinity delta series patient monitoring device
Hide thumbs Also See for Infinity Delta:
Table of Contents

Advertisement

Message
Possible Cause
PiCCO: Pod
PiCCO pod hardware failure.
Failure
PiCCO: Pod
The PiCCO pod is
Disconnected
disconnected.
VF8
D
/D
XL/K
ELTA
ELTA
Suggested Action
• Immediately remove sensor from patient.
• Check sensor; check pod and pod
connections. Replace sensor or pod, if
necessary.
• Plug and unplug the pod.
• Call your hospital's technical personnel or
DrägerService.
Check the cable and connection and replace, if
necessary.
APPA
S
M
TATUS
ESSAGES
26-29