The Unit Manager Menu
Menu Item
All Alarms OFF
2-18
Description
When the Code fixed key is pressed, the
following happens if the All Alarms OFF
setting set to Enabled:
• All alarms (audible and visual) are
disabled immediately and an All Alarms
OFF banner is displayed in the top
center of the monitor.
• A code timer is displayed in the bottom
left of the monitor.
• The Code Setup menu selection for
Alarm Volume OFF is disabled/
ghosted.
• If the Code function is disabled by
pressing the Code fixed key again, the
All Alarms OFF feature is deactivated
and the All Alarms OFF banner is
removed from the display.
When the Code fixed key is pressed, the
following happens if the All Alarms OFF
setting is set to Disabled:
• The Audio Volume setting is disabled
immediately and a speaker icon with a
slash mark is displayed in the top center
of the monitor.
• A code timer is displayed in the bottom
left of the monitor.
D
/D
ELTA
Available Settings
• Enabled
• Disabled (default)
XL/K
ELTA
APPA
VF8