4 Operation
4.5 Dentist element
F
E
92
4. Swivel the latch fastener B down onto the hose on the pump wheel.
Secure the latch fastener B with the locking mechanism A.
IMPORTANT
Flow rate of the peristaltic pump
The flow rate is set with the spring on the locking mechanism A. Do not
under any circumstances change the factory setting.
5. Push the short end of the hose with the cannula through the stopper
and into the NaCl bottle. The regulator in the hose clip E must be
H
completely open (regulating wheel in top position).
G
6. Run the long end of the hose alongside the corresponding motor
hose up to the contra-angle handpiece. Fasten the hose with clips F.
7. Attach the coupling G to the hose. Connect the thin silicone hose H
to the coupling G.
8. Connect the thin silicone hose H to the connectors on the contra-
angle handpiece. For details, see the Operating Instructions of the
contra-angle handpiece.
4.5.8.10

Setting the NaCl flow rate

An instrument is removed from its holder.
In the operating modes
Instrument program
touchscreen and NaCI is preselected as the coolant, see "Preselect
coolant" [ → 88].
In the
instrument is displayed.
1. In the
(> 2 s).
In the
pressed (> 2 s).
 The
2. Use the – and + keys to set the flow rate of the NaCl pump.
Simple/Advanced Start program
of the removed instrument is displayed on the
EasyMode Start program
, the
Simple/Advanced Start program
EasyMode Start program
NaCl
settings screen is displayed.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
, the
sub-screen
of the removed
NaCl
, keep the
key pressed
Preselect NaCl
, keep the
65 48 510 D3509
D3509.201.03.03.23 10.2016
key