Implantology

4.5.15.2
Contra-angle handpiece, implantology
CAUTION
Only the Sirona contra-angle handpiece IMPLANT 20:1 may be used for
the Implantology function.
Instruments from other manufacturers can lead to malfunctions. Outside
instruments may be improperly calibrated for implantology.
➢ Use only the IMPLANT 20:1 contra-angle handpiece from Sirona for
implantology.
➢ Check whether the reduction ratio on the touchscreen agrees with
the value specified on the contra-angle handpiece being used.
4.5.15.2.1
Calibrating the burr drive
Calibration is required at the start of treatment, every time the contra-
angle handpiece is changed and every time the contra-angle handpiece
is lubricated.
The contra-angle handpiece is automatically checked during calibration.
This includes a measurement of motor current at different speeds to
assess the properties of the system.
Calibration only with Sirona instruments
CAUTION
Use only Sirona instruments to guarantee correct calibration.
The
Implantology program
1. Attach the contra-angle handpiece that you would like to use for the
implantology treatment to the electric motor.
2. Insert the tool in the contra-angle handpiece. This is thus also taken
into account by the measurement.
3. Touch the
 The button flashes orange.
4. Hold down the foot pedal throughout the duration of the calibration.
 The
key permanently remains highlighted orange. Calibration is then
completed.
4.5.15.2.2
Setting the speed and torque
In the Implantology function, the speed and torque values of the contra-
angle handpiece, and not those of the motor, are specified. The control
electronics of the burr drive calculate the motor control based on the
specified gear reduction and the speed and torque values.
Implantology program
The
➢ Use the – and + keys to set the speed and torque of the contra-angle
handpiece. You can also hold down the keys for this purpose.
 The selected speed is displayed in the first line in rpm
(revolutions per minute). The torque is displayed in the second
line in Ncm (newton centimeters).
is displayed on the touchscreen.
Cal
key on the touchscreen.
Cal
key continues to flash. If the burr drive is calibrated, the
is displayed on the touchscreen.
4 Operation