Implantology/endodontics treatments with motor

4.5.10.4
4.5.10.4.1
Electronic torque limitation
If your dental treatment unit is equipped with the implantology/
endodontics software option, please observe the instructions provided in
the chapter "Implantology and endodontic treatments" [ → 122]. If your
dental treatment unit is not equipped with this option, please note that you
do not have the option of electronic torque limitation.
Endodontics with motor screen
CAUTION
No electronic torque limitation is available in the motor screen.
Root canal files can easily break during operation without torque
limitation.
➢ Never perform endodontic treatments without torque limitation. Use
an endodontic handpiece with mechanical torque limitation.
4.5.10.4.2
Endodontic treatments with the ApexLocator in the motor screen
If your dental treatment unit is equipped with the ApexLocator option but
does not have the treatment function, the Apex function in the Motor
program can be used for endodontic treatments.
Please note that the ApexLocator can only be accessed in the
screen
in the
IMPORTANT
ApexLocator, SIRONiTi APEX
SIRONiTi APEX
Only Sirona SIRONiTi APEX must be used to perform apex
measurement using a torque-limiting handpiece! The electrical
conductivity can only be ensured using this handpiece.
SIRONiTi APEX is not available in the USA.
The dental treatment unit is prepared for apex measurements using
the instrument, see section "Preparing to use the ApexLocator"
[ → 94].
Motor dialog
The
1. Set the appropriate motor speed according to the handpiece and the
root canal file used, see "Setting the speed" [ → 105].
2. You can make a setting to ensure that the motor stops automatically
at a preset distance from the apex. If you want to use the automatic
motor stop, this can be preset in the
touch the
 The
Simple/Advanced Start program
is displayed on the touchscreen.
Sub-dialog
key.
Motor
sub-dialog is displayed.
4 Operation
4.5 Dentist element
Motor
operating modes.
Motor
sub-dialog. To do this
109