4 Operation
4.5 Dentist element
126
IMPORTANT
Treatment, torque adjustment
Torque adjustment
The maximum adjustable torque depends on the system motor and the
speed settings.
Endodontics, speed and torque settings
CAUTION
Improperly selected speeds and torque values endanger the patient.
Treatment errors, e.g. jaw damage, may result from incorrect settings.
➢ Observe the manufacturer's instructions regarding tools and
implants.
4.5.15.2.3
Setting the direction of rotation
The direction of rotation can only be changed when the motor is idle.
Counterclockwise rotation is performed without torque limitation. The
torque setting keys are hidden when counterclockwise rotation is
selected.
Implantology program
The
➢ Touch the
or
Press the right button of the foot control.
 For counterclockwise rotation: The key
rotation
rotation arrow appears.
For clockwise rotation: The key
displayed in gray and the orange counterclockwise arrow is
hidden.
Tip: After starting the electric motor with the foot switch, an audible
warning signal sounds 6 times if counterclockwise rotation is activated.
4.5.15.2.4
Activating/deactivating and setting the NaCl flow
In order to supply the burr drive with a sterile saline solution during
implantological treatments, a peristaltic pump must be attached to the
dentist element; see "Preparing for use of NaCl saline solution" [ → 91].
The peristaltic pump can be switched on/off by touching the
the key is highlighted orange, the pump can be activated by stepping on
the foot control.
The NaCl flow rate set of the peristaltic pump is permanently displayed by
a bar in the third line of the touchscreen in the
The flow rate can be set by pressing and holding the
than 2 seconds). For details, see "Setting the NaCl flow rate" [ → 92].
is displayed on the touchscreen.
Counterclockwise rotation
is highlighted orange and an orange counterclockwise
Counterclockwise rotation
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
key on the touchscreen.
Counterclockwise
NaCl
Implantology
program.
NaCl
key (for more
65 48 510 D3509
D3509.201.03.03.23 10.2016