Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
65 48 510 D3509
D3509.201.03.03.23 10.2016
5 Care, cleaning and maintenance by the practice team
1. Switch off the dental treatment unit via the user interface, not via the
power switch on the base of the patient chair. Make sure that the
dental treatment unit is switched off for at least 24 hours, but not
longer than 3 days (maximum sanitation time).
2. Switch the dental treatment unit back on after the reaction time.
Rinsing DENTOSEPT P out of the mixing tank
Once it has reacted, the remaining DENTOSEPT P must be rinsed out of
the mixing tank using the SPRAYVIT.
1. Insert the SPRAYVIT valve body on the dentist element into the water
unit adapter as described above. Wait until no more water escapes
from the SPRAYVIT.
 The warning triangle with error code 10 ("Flow at the SPRAYVIT
M on the dentist element too low") appears again in the status bar
of the touchscreen. The remaining DENTOSEPT P is now rinsed
out.
2. Pull the SPRAYVIT valve body out of the water unit and reattach the
SPRAYVIT sleeve. Return the SPRAYVIT to the instrument holder.
Purging water lines with water
The DENTOSEPT P waterline treatment solution is rinsed out of the
instrument hoses, the SPRAYVIT hoses, and tumbler filler with water.
1. Mix distilled water with the DENTOSEPT P waterline treatment
solution at a ratio of 100:1 (1 liter of water, 10 ml DENTOSEPT P)
and fill this into the tank for waterline treatment solution of the water
unit. Then wait for approx. 2 minutes.
2. Hold the instruments over the cuspidor or over a watertight container
with sufficient capacity again and thoroughly rinse all the instruments
for approximately 30 seconds.
3. Empty the rinsing tumbler and place the empty tumbler again below
the tumbler outlet.
4. If your dental treatment unit is equipped with a cuspidor, press the
tumbler filler three times.
 The manual sanitation process is finished. The dental treatment unit
is again ready for operation.
5.7 Sanitation
259