10 Overview of all function keys
10.4 Treatment program
278
Direct starter/speed foot control
Direct starter (highlighted gray): Switches the instrument on with the set
speed or intensity
Speed foot control (highlighted orange): Regulates the instrument to the
set speed or intensity after setting the foot switch to maximum.
Setting the number of work steps
Up to eight implantology work steps can be saved.
10.4.3
Endodontics
Calibrating the burr drive
A calibration must be performed each time you change or lubricate the
contra-angle handpiece.
The contra-angle handpiece is automatically checked during calibration.
This includes a measurement of motor current at different speeds to
assess the properties of the system.
Direction of rotation
Switches counterclockwise rotation ON/OFF
Insert file
Opens a screen in which individual files, even from other file systems, can
be inserted in the existing sequence.
Creating user-defined files
Creation of individual user-defined files. Once a name has been entered,
the standard speed and torque values of the new file must be set
according to the manufacturer's specifications.
Select file system
Each time the key is pressed, the next file system in the list is displayed
on the left side of the touchscreen.
Opening the sub-screen
Used to access additional subordinate functions; see the following
function keys:
Setting the automatic motor stop of the ApexLocator
Apex Stop
When the
automatic motor stop can be switched off or set to four different levels. If
the distance is 0, the motor does not stop until it reaches the physiological
apex. Please note that the distance values are not a metric length
measurement!
The preset motor stop position is displayed as a black triangle to the right
of the distance display under the text "Stop".
Apex distance acoustic signals
Switching the apex distance signal tones on or off. If the automatic motor
stop is switched off, the intervals between the beeps vary according to the
measured distance from the apex. If this function is switched on, the
beeps vary depending on the measured distance to the preset motor stop
position.
Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
key is pressed, the – and + keys are displayed. The
D3509.201.03.03.23 10.2016