Canceling the sanitizing process
Sanitation can be canceled during some sanitation phases, e.g. if the
dental treatment unit must be switched back on before the 24 hours have
Stop
expired. The
The sanitizing process has started.
Stop
➢ Touch the
 The sanitation process is canceled. If the water lines have not yet
been treated with DENTOSEPT P waterline treatment solution
(prior to sanitation phase 3), the dental treatment unit
automatically proceeds directly to sanitation phase 6. If the
sanitation process is canceled during the reaction time, the water
lines are initially rinsed with water (sanitation phase 5).
5.7.2

Display sanitation report

Sanitation processes with interactive guidance, with a reaction time of at
Sanitation processes with interactive guidance
least 24 hours, are logged by the dental treatment unit. The
corresponding log reports can be displayed on the touchscreen.
Sanitation program
The
Sanitation program
➢ Press the
 The sanitation report is displayed.
The sanitation report contains the following data:
Column
Display
6 instrument holders on the dentist element
1
Sequential number
2
Status of sanitized instruments in dentist element, holder 1
to 6
3
Status of sanitized instruments in assistant element, holder
1 to 4
4
Date of sanitation
5
Time
The status of column 3 and 4 can assume the following values:
+ = successful sanitation
- = Sanitation was not completed (e.g., interruption)
5 Care, cleaning and maintenance by the practice team
key shows whether cancellation is possible.
key on the touchscreen.
is displayed on the touchscreen.
key to display the log.
5.7 Sanitation